| 48T | ¥1680.00 |
| 96T | ¥2400.00 |
| 96T*5 | ¥10200.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 2.389 | 2.307 |
| 250.00 | 1.593 | 1.511 |
| 125.00 | 1.213 | 1.131 |
| 62.50 | 0.882 | 0.800 |
| 31.25 | 0.585 | 0.503 |
| 15.63 | 0.305 | 0.223 |
| 7.82 | 0.232 | 0.150 |
| 0.00 | 0.082 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 86-94% | 90% |
| EDTA plasma(n=5) | 77-84% | 81% |
| Heparin plasma(n=5) | 84-90% | 87% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 80-92% | 96-105% | 84-95% | 89-102% |
| EDTA plasma(n=5) | 87-98% | 86-95% | 96-107% | 96-107% |
| Heparin plasma(n=5) | 83-95% | 84-97% | 87-98% | 84-96% |